MODEL |
AG9323 |
Output Channel 1 & 2 |
Voltage |
0 ~ 32V |
Current |
0 ~ 3A |
OVP |
0.1 ~34V |
OCP |
Enable/Disable |
Output Channel 3 |
Voltage |
Range 1 : 0 ~ 32V, Range 2 : 0 ~ 6V |
Current |
Range 1 : 0 ~ 1A, Range 2 : 0 ~ 5A |
OVP |
0.1 ~ 35V |
OCP |
Enable/Disable |
Line Regulation |
Voltage |
≦0.01%+3 mV |
Current |
≦0.06%+3mA (+6mA rating current >3A) |
Load Regulation |
Voltage |
≦0.01%+3 mV |
Current |
≦0.06%+3mA |
Ripple & Noise |
Voltage |
≦1mVrms |
Current |
≦3mArms |
Program Resolution |
Voltage |
10mV |
Current |
1mA |
Program Accuracy |
Voltage |
≦0.05+10 mV |
Current |
≦0.08%+5mA |
Program Temperature |
Voltage |
≦100ppm/degree C +3mV ( 0 ~ 40 degree C) |
Current |
≦150ppm/degree C +3mA ( 0 ~ 40 degree C) |
Readback Resolution |
Voltage |
10Mv |
Current |
1mA |
Readback Accuracy |
Voltage |
≦0.05%+10 mV |
Current |
≦0.08%+5mA |
Response Time |
Voltage Up & Down ≦100ms |
Readback Temperature |
Voltage |
≦100ppm/degree C +3mV ( 0 ~ 40degree C) |
Current |
≦150ppm/degree C +3mA ( 0 ~ 40 degree C) |
Output Drift |
Voltage |
≦100ppm/degree C +10mV ( 0 ~ 40 degree C) |
Current |
≦150ppm/degree C +10mA ( 0 ~ 40 degree C) |
Tracking Mode |
Tracking Error |
≦0.1%+20mV |
Memory |
Save & Recall |
1 to 50 points |
Timer Setting |
Setting Time |
0.1 to 9999.9 sec |
Resolution |
0.1 sec |
Function |
For output working loop (Auto Step Running) |
Interface |
RS-232 |
9600,N,8,1 |
Power Source |
ACV 100/120/220/240V, ±10%, 60/50Hz |
Temperature Rating |
Operating |
0 ~ 40 degree C |
Storage |
-10 ~ 70 degree C |
Safety & EMC |
CE (include EMC & LVD) |
General |
Accessories |
Test Lead x3, Power Cord x1, Operation Manual x1 |
Dimensions |
270(W) x 156(H) x 462(D) mm |
Weight |
16 kg |