The product have designed specifically to operate
with strain gauge sensors such as Fast-
measurement of Load cell, Pressure Transducer,
LVDT, etc. The setup sequence is simple and
quick using four push buttons only.
-
High resolving power (16bit A/D)
-
High presision (more than 0.012% F.S) and
stability
-
Quick sampling speed (100times/sec)
-
Built-in Shunt calibration switch
-
Watch dog
-
Small size(96*48*128mm),
Cutting size : 92*45mm
-
Weight : about 700g
|
|
|
|
DN15W-Various Functions
-
Calibration of actual load and sensor output value
-
Peak hold and Sample hold
-
Auto zero
DN10W-Various Functions
-
Calibraion of actual load and sensor output value
-
Peak hold and Sample hold
-
Display part and analogue zero by auto zero (simultaneous zero of both digital display and
analogue output)
-
Upper and lower limit comparison output
- Comparison output :
ON in case of Hi-set value≤input value,
ON in case of LO-set value≥input value
- Contact capacity : AC 250V 0.5A, DC 30V 0.5A
- Relay life time : more than 200,000,000 times
-
Analog output : 0~10V or 4~20mA
-
Option : RS232C (OP-01), RS485(OP-02)
|
|
|
Specifications
|
Accuracy
|
Application sensor
|
Strain guage sensor (350Ω or 120Ω)
|
Sensor authorized voltage
(the number of Loadcell connection)
|
DC 5V, Loadcell(350Ω) available 4
|
Zero adjustment range
|
100% F·S (Auto Zero)
|
Input signal
|
0~3㎷/V
|
Display
|
-19999~99999
|
Display speed
|
Appox 30times/sec(Adjustment available by buff setting)
|
Character height
|
7Segment 14㎜
|
Decimal point
|
Random setting available
|
Sampling speed
|
Max. 100times/sec
|
Zero variation
|
Within 0.5㎶/℃
|
Sensitivity variation
|
Within 0.01%/℃
|
Voltage output
|
DC 0~10V /
|
Usage temperature range
|
-10~60℃
|
Humidity
|
less than 80% RH (no dew condensation)
|
Power source used
|
220V 50/60㎐ (110V available by internal operation)
|
Relay output
|
- Upper and lower limit comparison output
-4 Relay output
-
Contact capacity : AC 250V 0.5A, DC 30V 0.5A
-
Relay life time : more than 20,000,000 times
|
Option
|
RS232C(OP-01), RS485(OP-02), DAC ANALOG OUT(OP-03) Input 5㎷/V : LVDT(OP-04)
4~20㎃ (OP-05)
|
|
|
|
Specifications
|
Accuracy
|
Application sensor
|
Strain guage sensor (350Ω or 120Ω)
|
Sensor authorized voltage
(the number of Loadcell connection)
|
DC 5V, Loadcell(350Ω) available 4
DC 10V, Loadcell(120Ω) available 2
|
Zero adjustment range
|
100% F·S (Auto Zero)
|
Input signal
|
1.0, 1.5, 2, 3㎷/V (Dip switch)
|
Display
|
-19999~ + 32767
|
Display speed
|
Appox 30times/sec(Adjustment available by buff setting)
|
Character height
|
7Segment 14㎜
|
Decimal point
|
Random setting available
|
Sampling speed
|
Max. 100times/sec
|
Zero variation
|
Within 0.5㎶/℃
|
Sensitivity variation
|
Within 0.01%/℃
|
Usage temperature range
|
-10~60℃
|
Humidity
|
less than 80% RH (no dew condensation)
|
Power source used
|
220V 50/60㎐ (110V available by internal operation)
|
|
|
|
|